1. Introduction
2152 Model ROSS-Miles Tester (Improved ROSS-Miles Method) meets the technical requirements of GB/T7462-94; national standard. It is suitable for the determination of the foaming power of surfactants. (Surfactants such as: dodecyl dimethyl betaine (BS-12), dodecyl dimethyl amine oxide (OA-12), coconut oil acylated propyl dimethyl betaine (CAB-35), coconut oil acylated propyl dimethyl amine oxide, fatty alcohol polyoxyethylene ether phosphate potassium salt, etc.)
1. The Roche foam meter must be installed vertically, otherwise the liquid surface will be uneven and the reading will be inaccurate.
2. The test solution should be preheated to about 41.5ºC before being placed in the dropper. The temperature during formal operation after injection should be 40ºC±0.5, otherwise the temperature will have a great impact on the data.
3. The foam activity of some solutions is very unstable. After a few minutes, the foam surface will break and become an uneven surface. At this time, the height reading can only be an estimated average number.
2. The test solution should be preheated to about 41.5ºC before being placed in the dropper. The temperature during formal operation after injection should be 40ºC±0.5, otherwise the temperature will have a great impact on the data.
3. The foam activity of some solutions is very unstable. After a few minutes, the foam surface will break and become an uneven surface. At this time, the height reading can only be an estimated average number.
2. Specification
1 |
Machine name
|
ROSS-Miles Surfactant Analysis Meter
|
2 |
Material
|
B40 glass
|
3 |
Dropping tube capacity
|
200ml
|
4 |
Dropping tube total length
|
300mm
|
5 |
Inner diameter of graduated tube
|
¢50mm
|
6 |
Graduated tube total length
|
1100mm
|
7 |
Frame support
|
SUS
|